Product details & mandatory information
Calibration of the Urisys 1100
After the Urisys system has already been calibrated at the factory before delivery, the process must be repeated when the system is first put into operation and then every 7 days using the Control-Test M strips. The Control-Test M consists of a standard grey plastic strip with a defined, constant remission. The calibration compensates for the ageing processes of the optics and the grey reference field in the slide. If there are significant deviations, for example due to a dirty reference field or the light intensity of a defective LED is too low, an error message appears.
The recalibration is automatically requested by the Urisys 1100 system on a weekly basis.
Ideally suited for
- Hospitals
- Doctors' surgeries
- Emergency admissions
Advantages
- Automatic request: The recalibration is automatically requested by the Urisys 1100 system on a weekly basis.
- Automatic saving: If the calibration is valid, the device automatically saves the result including the date and time.
Disadvantages
- High sensitivity: When removing the calibration strip, the fields must not be touched with fingers or urine, otherwise the strip will become invalid.